Configuration Information
3500/42 Operation and Maintenance
12
Prox/Velom Internal Barrier I/O
The transducer field wiring is connected directly to the Proximitor/Seismic
Monitor Internal Barrier I/O Module. Note that selecting the Prox/Velom
Internal Barrier I/O option will disable certain transducer type options.
Velom Internal Barrier I/O
The transducer field wiring is connected directly to the Proximitor/Seismic
Monitor Internal Barrier I/O Module. Note that selecting the Velom Internal
Barrier I/O option will disable certain transducer type options.
TMR I/O
Used when three identical adjacent monitors and three TMR I/O Modules are
installed for a TMR application. Both the discrete and bussed configurations
use the same external I/O modules but are wired differently as per the
following paragraphs.
TMR I/O (Discrete)
This option is used when redundant transducers and field wiring are required.
A set of twelve transducers are used to provide input signals to three identical
adjacent monitors. Each transducer is connected to an External Termination
Block and then routed to the Proximitor/Seismic TMR I/O Module using a 25-
pin cable. The recorder field wiring is connected to an External Termination
Block and then routed from the External Termination Block to the
Proximitor/Seismic TMR I/O Module through a 9-pin cable.
TMR I/O (Bussed)
This option is used when redundant transducers and field wiring are not
required. A single set of four transducers are sent to three identical adjacent
monitors. Each transducer is connected to a Bussed External Termination
Block and then the Bussed External Termination Block is connected to the
Proximitor/Seismic TMR I/O Modules using three 25-pin cables. The
recorder field wiring is connected to an External Termination Block and then
routed from the External Termination Block to the Proximitor/Seismic TMR
I/O Module through a 9-pin cable.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com