GUI Reference
R&S
®
ZNA
279
User Manual 1178.6462.02 ─ 12
6.2.7.1
Controls on the Gain Compression Tab
Amplifier Compression/Setup Frequency Converting DUT
for gain compression measurements:
●
Chapter 6.2.7.2, "Amplifier Compression Setup Dialog"
on page 279 for non-
frequency converting DUTs
●
Chapter 6.2.7.3, "Converter Compression Setup Dialog"
on page 281 for fre-
quency-converting DUTs
Compression Point Power In/Compression Point Power Out
Selects the input or output value of the <x
dB> compression point as the measured
quantity, where <x
dB> is the specified
Remote command:
'
CALCulate<Ch>:PARameter:MEASure
'<TraceName>',
'CmpPtPin'|'CmpPtPout'
CALCulate<Ch>:PARameter:SDEFine
'<TraceName>',
'CmpPtPin'|'CmpPtPout'
Compression Value
With a "Compression Value" of
x
, the gain compression measurement searches for the
x
dB compression points of the power sweeps.
The
x
dB compression point is defined as the stimulus level where the response value
has dropped by
x
dB compared to the response value at
.
Remote command:
[SENSe<Ch>:]FREQuency:COMPression:POINt
6.2.7.2
Amplifier Compression Setup Dialog
The "Amplifier Compression" setup dialog is an instance of the
. It allows you to set up one or more channels for compression point over fre-
quency measurements on non-frequency converting DUTs.
Access:
> "S-Params" > "Gain Compression" > "Amplifier Compres-
sion"