GUI Reference
R&S
®
ZNA
455
User Manual 1178.6462.02 ─ 12
Tip:
If the characterization wizard is password-protected, the "Characterization" button
is unavailable. Use this functionality to prevent inadvertent activation of inappropriate
characterizations. See
See also
Chapter 5.5.5.3, "Characterization of Calibration Units"
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
Query further cal unit properties:
SYSTem:COMMunicate:RDEVice:AKAL:DATE?
SYSTem:COMMunicate:RDEVice:AKAL:FRANge?
SYSTem:COMMunicate:RDEVice:AKAL:PORTs?
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Calibration ← Cal. Settings
Displays the calibration type.
Tap the table cell to open the "Define Calibration" dialog that lets you choose calibrated
ports and type of the calibration definition.
Figure 6-18: Define Calibration dialog: manual calibration
Cal Softtool