GUI Reference
R&S
®
ZNA
440
User Manual 1178.6462.02 ─ 12
Background information
Refer to the following sections:
●
●
Chapter 5.5.5, "Automatic Calibration"
●
Chapter 5.5.6, "Scalar Power Calibration"
●
●
6.11.1.1
Measurement Modes and Default Calibration Setup
, or the
for the first time, the analyzer determines the preferred calibration setup
based on:
●
the channel's (or channels') measurement mode (or modes)
●
the channel's traces
●
whether a power meter is connected
The channel's (or channels') calibration setup is either written to the recall set automat-
ically, or you are asked to select between several reasonable setups (non-"Quick Start"
actions only).
Figure 6-16: Select Calibration Properties dialog: scalar mixer measurements
For the standard mode a calibration unit is used, if connected.
Measurement
mode
Traces
Power meter
Calibration setup
Standard
only one-port
–
OSM
✓
POSM
at least one two-
port
–
UOSM
Cal Softtool