Command Reference
R&S
®
ZNA
1353
User Manual 1178.6462.02 ─ 12
FDATa
Formatted trace data, according to the selected trace format (
). 1 value per trace point for Cartesian diagrams, 2 values for polar dia-
grams.
SDATa
Unformatted trace data: Real and imaginary part of each measurement point. 2
values per trace point irrespective of the selected trace format. The trace mathe-
matics is not taken into account.
MDATa
Unformatted trace data (see
SDATa
) after evaluation of the trace mathematics.
INPut<PhyPt>:ATTenuation
<Attenuation>
Sets the attenuation for the received waves. This command is available if at least one
of the
The generated wave is attenuated via
SOURce<Ch>:POWer<PhyPt>:ATTenuation
.
For redefined physical ports (see
measurement receiver (b-wave) is significant. E.g. if a receover step attenuator is
installed for physical port 1, then an attenuation factor can be applied to the (redefined)
port receiving b1.
Note
:
●
INPut<PhyPt>:ATTenuation
is not channel-specific; the value is valid for
all
channels
. Use
specific attenuation value.
Suffix:
<PhyPt>
.
Physical port number; if unspecified the numeric suffix is set to 1
Parameters:
<Attenuation>
Attenuation factor for the received wave.
Range:
0 dB, 10 dB, 20 dB, 30 dB. UP and DOWN incre-
ment/decrement the attenuation in 10 dB steps.
The analyzer rounds any entered value below the
maximum attenuation to the closest step.
*RST:
0 dB
Default unit: dB
Example:
INP2:ATT 10
Set the step attenuator for the wave received at port 2 and for all
channels to 10 dB. The waves at the other test ports are not
affected.
SENS1:POW:ATT? BREC
Query the receiver step attenuator setting at port 2 and for chan-
nel no. 1. The response is
10
.
INSTrument[:SELect]
<Channel>
Selects a channel
<Ch>
as active channel. To select a channel number > 4 use the
generalized command
R&S ZVR/ZVABT Compatible Commands