Command Reference
R&S
®
ZNA
1258
User Manual 1178.6462.02 ─ 12
<PhyPt>
Calibrated port number. This parameter is ignored; the setting is
valid for all sources.
Parameters:
<Boolean>
ON - other sources not necessarily switched off.
OFF - other sources switched off during the calibration sweep.
*RST:
ON
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:PPOWer
<RecCalPower>
Defines the source power which the R&S
ZNA uses to perform the first calibration
sweep of the source power calibration ("Reference Receiver Cal Power"). The power
value is ignored if the R&S
ZNA is set to use the port power result (
POWer<PhyPt>:CORRection:PSELect
CPOW
).
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
parameter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parame-
ter is valid for all sources.
Parameters:
<RecCalPower>
Reference receiver cal power
Range:
Depending on source power range of the analyzer
and the power range of the external power meter.
*RST:
0 dBm
Default unit: dBm
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation:
See
"Reference Receiver Cal Power"
SOURce<Ch>:POWer<PhyPt>:CORRection:PSELect
<CalPower>
Qualifies how to define the source power which the R&S
ZNA uses to perform the first
calibration sweep of the source power calibration ("Reference Receiver Cal Power").
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
parameter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parame-
ter is valid for all sources.
SCPI Command Reference