<CorrectionData>
Correction data set (one complex number per sweep point) to be
transferred to the analyzer either in ASCII or block data format,
depending on the current
tion data set is assigned to the specified channel, error term,
source and load port.
This parameter must not be used for queries.
Example:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE:
The different calibration types of the analyzer provide the following error terms:
Calibration type
Parameters in
[SENSe<Ch>:]CORRection:
COLLect:METHod:DEFine
Available error terms (depending on port num-
bers)
One-port normalization (reflection)
using an open or a short standard
REFL, RSHort
'REFLTRACK'
Full one port ("Refl OSM")
FOPort
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK'
Two-port normalization
FRTRans
'TRANSTRACK'
One path two port
OPTPort
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'TRANSTRACK'
TOSM
TOSM
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'LOADMATCH', 'TRANSTRACK'
TOM, TSM, TRM, TRL, TNA
TOM | TSM | TRM | TRL | TNA
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'LOADMATCH', 'TRANSTRACK' (for reading and
writing)
'G11' ... 'G22' and 'H11', 'H12', 'H22' (for reading
only; the 'H21' matrix elements are normalized to
1)
8.3.14.7
[SENSe:]CORRection:CKIT...
The
[SENSe:]CORRection:CKIT...
commands deal with calibration kits and cal kit
data. The calibration kits are distinguished by their names (<CalkitName>), the optional
labels (<label>) can be used to carry information about the calibration standard.
In order to handle several identical calibration kits with different serial numbers use the
commands of
Chapter 8.3.14.8, "[SENSe:]CORRection:CKIT... with Labels"
[SENSe:]CORRection:CKIT:CATalog?
............................................................................ 1068
[SENSe:]CORRection:CKIT:DELete
............................................................................... 1068
.............................................................................. 1068
................................................................................. 1069
[SENSe:]CORRection:CKIT:SELect
............................................................................... 1069
[SENSe:]CORRection:CKIT:STANdard:CATalog?
............................................................ 1070
[SENSe:]CORRection:CKIT:STANdard:DATA?
................................................................ 1070
[SENSe:]CORRection:CKIT:<ConnType>:SELect
............................................................ 1071
[SENSe:]CORRection:CKIT:<StandardType>
.................................................................. 1072
SCPI Command Reference