![Rohde & Schwarz 1332.4500.22 User Manual Download Page 530](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782530.webp)
GUI Reference
R&S
®
ZNA
530
User Manual 1178.6462.02 ─ 12
Parameter
Lin/Log Freq
Sweep
Power Sweep
CW Mode
Sweep
Time Sweep
x
x
x
x
x
Other params
x
x
x
Enter & Display
The "Enter & Display" combo boxes in the "Frequency Sweep" and "Power Sweep"
sections of the "Stimulus" dialog control how frequency and power values are entered
and displayed.
●
channel base frequency fb/power
●
frequency/power at logical port i
●
source frequency/power at logical port i
●
receiver frequency at logical port i
Note that setting "Enter & Display" to a particular port, source, or receiver is only possi-
ble if option R&S
ZNA-K4 is available. Furthermore, a particular port, source, or
receiver can only be selected, if the user-defined linear function for calculating the
respective frequency/power is not constant.
6.12.2.3
Define Parallel Measurement Dialog
With a 4-port R&S
ZNA, you can measure multiple DUTs (or multiple paths of a single
DUT) in parallel. The "Define Parallel Measurement" dialog allows you to configure
these DUTs and to declare the connections between physical VNA ports and DUT
ports. Furthermore it provides convenient access to port-related settings.
Related information
See
Chapter 5.1.4.2, "Parallel Measurements on Multiple DUTs"
The "Parallel Measurement" is automatically configured when "Enabled" is selected, as
shown in
.
A R&S
ZNA with two internal sources allows parallel measurements with "Frequency
Offset", which can be configured in the corresponding tab on the right.
Channel Config Softtool