Command Reference
R&S
®
ZNA
1213
User Manual 1178.6462.02 ─ 12
Manual operation:
See
[SENSe<Ch>:]SWEep:TYPE
<Format>
Selects the sweep type, i.e. the sweep variable (frequency/power/time) and the posi-
tion of the sweep points across the sweep range.
Suffix:
<Ch>
.
Channel number.
Parameters:
<Format>
LINear | LOGarithmic | POWer | CW | POINt | SEGMent |
PULSe | IAMPlitude | IPHase
LINear
Linear frequency sweep at constant source power
(
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:
(
[SENSe<Ch>:]FREQuency:...
) is swept in equidistant steps
over the frequency range. In a Cartesian diagram, the x-axis is a
linear frequency axis.
LOGarithmic
Logarithmic frequency sweep. The frequency is swept in equi-
distant steps on a logarithmic scale. In a Cartesian diagram, the
x-axis is a logarithmic frequency axis.
POWer
Power sweep. The measurement is performed at constant fre-
quency (
SOURce<Ch>:FREQuency<PhyPt>:FIXed
variable generator power that is swept in linear, equidistant
steps over a continuous range
(
SOURce<Ch>:POWer<PhyPt>:STARt
). In a Cartesian diagram, the x-axis is a
dB-linear power axis.
CW
Time sweep. The measurement is performed at constant fre-
quency (
SOURce<Ch>:FREQuency<PhyPt>:FIXed
SOURce<Ch>:POWer<PhyPt>[:LEVel][:
) and repeated over a specified
period of time at constant time intervals.
POINt
CW mode sweep, time sweep triggered according to the current
trigger settings.
SEGMent - segmented frequency sweep. The sweep range is
composed of several continuous frequency ranges or single fre-
quency points defined by means of the commands in the
[SENSe<Ch>:]SEGMent<Seg>...
subsystem.
PULSe
Pulse profile mode, configured via
SENSe:PULSe...
SCPI Command Reference