Firmware Version 2.10
R&S
®
ZNA
8
User Manual 1178.6462.02 ─ 12
–
Overall and individual test results clearly highlighted ("PASS": green, "FAIL":
red)
●
Measurement uncertainty analysis (R&S
ZNA-K50): zip archives created using
Export Data to METAS VNA Tools
can now be directly unpacked to the user data
folder of a METAS VNA Tools installation for further offline analysis
●
Source coherence mode (R&S
ZNA-K6): direct phase control, without applying port
match and calibration
See
"Use Port Match and Calibration"
●
If compatible with the overall converter setup, the source providing the LO IN signal
to a frequency converter is automatically configured as permanent ("Source Gen")
●
Segmented sweeps: changes take effect faster (less "prepare sweep" phases)
●
Improved access to extended [Preset] functionality via menu bar
●
Manual ALC control loop configuration: new parameters "Control Time" and "Auto
Settling"
See
Solved issues
●
Factory correction for Direct Access (B16) inputs was not applied in SA mode
●
Frequency-converting measurements with Converter LO used as LO provider: the
"Power Result" displayed in the "Port Settings" dialog was not correct
●
Frequency-converting measurements with external generator used as LO provider:
excessive retrace time
●
Some timing/synchronization problems with external generator R&S
SGS100A
●
Misleading calibration instructions on "DUT Centric Wizard" start page
●
Calibration setup dialog:
–
redundant warning messages
–
sometimes a warning message did not disappear after the underlying reason
has been eliminated
●
IMD measurements:
–
Inconsistent warning messages during default calibration
–
Misspelled measured quantity for Through measurement during calibration
(now labeled LTO for "Lower Tone at DUT Out")
●
Deleting a trace by dragging it to the recycle bin sometimes did not work
●
Configured "Reference Receiver Cal Power" value not used for SMARTerCal
●
Formula-defined limit lines: formula was not applied if entered directly in
"Response" dialog
●
Impedance normalization failed for some special balanced port configurations
●
*.calkit
files created with R&S
ZNA and containing snp data were not compati-
ble with R&S
ZNB/BT/C/D firmware versions < 3.0 and R&S
ZNL/LE firmware ver-
sions < 1.30