![Rohde & Schwarz 1332.4500.22 User Manual Download Page 918](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782918.webp)
Command Reference
R&S
®
ZNA
918
User Manual 1178.6462.02 ─ 12
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
Parameters:
<Boolean>
ON | OFF - embedding active or inactive
*RST:
OFF
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:C<Cmp>
<CircuitModel>[, <Capacitance>]
Specifies the capacitance value C in the "Shunt L, Shunt C" lumped element model for
differential match embedding.
In the query form, the <Capacitance> parameter must be omitted.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
<Cmp>
Must be omitted or set to 1.
Parameters:
<CircuitModel>
SHLC
Currently only the "Shunt L, Shunt C" lumped element model is
supported
<Capacitance>
Range:
-1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST:
1 pF (1E-12 F)
Default unit: F
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:DATA
<Interchange>, <SParamTrcs>
Defines a
network for a balanced port based on the
given S-Parameter traces.
Use
MMEMory:LOAD:VNETworks<Ch>:DIFFerential:EMBedding<LogPt>
to
load circuit data from a Touchstone file located at the R&S
ZNA's file system instead.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number of a balanced port
Setting parameters:
<Interchange>
FPORts | IPORts | SGATes
SCPI Command Reference