Command Reference
R&S
®
ZNA
803
User Manual 1178.6462.02 ─ 12
Manual operation:
See
CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP
<String>
Tells the ISD tool which ports (in the measured DUT + Test Fixture file) shall be skip-
ped when the tool is run.
Equivalent to using the
ports_to_skip
batch mode parameter of the ISD tool with a
list of (positive) port numbers.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<String>
The port numbers, separated by blanks (e.g. '1 3 4').
Manual operation:
See
CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP:LEFT
CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP:NONE
CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP:RIGHt
Tells the ISD tool which ports (in the measured DUT + Test Fixture file) shall be skip-
ped when the tool is run.
●
...:LEFT
: skip the ports on the left (according to the port order specified using
CALCulate:FMODel:ISD<Ph_pt>:PORT:ORDer
●
...:NONE
: do not skip any ports
●
...:RIGHt
: skip the ports on the right (according to the port order specified using
CALCulate:FMODel:ISD<Ph_pt>:PORT:ORDer
)
Equivalent to setting the
ports_to_skip
batch mode parameter of the ISD tool to -1,
0 or -2, respectively.
Use
CALCulate:FMODel:ISD<Ph_pt>:PORT:SKIP
to define an arbitrary set of
ports to be skipped.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage:
Event
Manual operation:
See
CALCulate:FMODel:ISD<Ph_pt>:RUN:RUN
Runs the ISD tool.
Before executing this command, make sure that
●
the test coupon measurement (using
CALCulate:FMODel:ISD<Ph_pt>:COUPon:
otherwise) finished successfully, or the test coupon data were