Programming Examples
R&S
®
ZNA
1404
User Manual 1178.6462.02 ─ 12
CONFigure:CHANnel3 ON // channel 3
CALCulate3:PARameter:SDEFine 'LO_Leak','S13'
DISPlay:WINDow1:TRACe5:FEED 'LO_Leak'
CALCulate3:PARameter:SDEFine 'LO_Thru','S23'
DISPlay:WINDow1:TRACe6:FEED 'LO_Thru'
//channel names
CONFigure:CHANnel1:NAME 'Ch_M'
CONFigure:CHANnel2:NAME 'CH_RF'
CONFigure:CHANnel3:NAME 'CH_LO'
// Adjust channels for measurement
//channel 1 (arbitrary) remains unchanged
//channel 2 (base frequency), reset arbitrary mode
SENSe2:FREQuency1:CONVersion FUNDamental
// trace RF_Isol in channel Ch_RF is measured with active LO!
SOURce2:FREQuency3:CONVersion:ARBitrary:IFRequency 1, 1, 3 GHz, FIXed
SOURce2:POWer3:PERManent:STATe ON
//channel 3 (LO freqeuncy), reset arbitrary and set LO-frequencies
SENSe3:FREQuency1:CONVersion FUNDamental
//LO-fix : Start=Stop, sweep-points = 1
SENSe3:SWEep:POINts 1
SENSe3:FREQuency1:STARt 200 MHz
SENSe3:FREQuency1:STOP 200 MHz
9.2.7
Fixture Modeling
We create and apply a fixture model for single-ended deembedding at ports 1 and 2
using the ISD tool.
Prepare the channel
*RST
:SENSe1:FREQuency:STARt MINimum
:SENSe1:FREQuency:STOP MAXimum
Perform a full two-port calibration at ports 1 and 2
For an example, see
Measure the test coupon
We are using a single-ended (2-port) "Symmetric 2x Thru" coupon (balanced). Connect
it to ports 1 and 2.
Condensed Programming Examples