![Rohde & Schwarz 1332.4500.22 User Manual Download Page 697](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782697.webp)
GUI Reference
R&S
®
ZNA
697
User Manual 1178.6462.02 ─ 12
If an instrument is connected to the R&S
ZNA, the entries in the [Driver for New
Instrument] panel can be auto-detected for the specified interface type and
address.
●
"Identify" sends an identification query ("IDN?") to the specified device address in
order to identify the type and serial number of the connected generator and select
an appropriate driver file. Generator driver files (
*.gen
) are stored in the
Resources\ExtDev
subdirectory of the analyzer's program directory.
Table 6-10: Interface types for external generators and address formats
Physical
interface
(connector)
Interface
(protocol)
Address
Remarks
LAN
VXI-11
<IpAddress>
e.g.
127.0.0.0
Full VISA resource string:
TCPIP[board]::<Address>[::INSTR]
SOCKET
<IpAddress>::<PortNo>
e.g.
127.0.0.0::50000
LAN connection with pure TCP/IP protocol;
refer to your VISA user documentation.
GPIB
GPIB0 ...
GPIB9
<Address>
e.g.
20
Full VISA resource string:
GPIB[board]::<Address>[::INSTR]
USB
USB-VISA
<ManID>::<ProdID>::<SerialNo>
e.g.
0x0AAD::0x0047::100098
2733 (0x0AAD) is the manufacturer ID of
Rohde & Schwarz.
71 (0x0047) is an example for a R&S prod-
uct ID (R&S SMF100A).
The serial number is device-specific.
LAN or USB Other
Interface-specific, e.g. for SOCKET:
TCPIP0::<IpAddress>::<PortNo>::SOCKET
For USB-VISA:
USB0::<ManID>::<ProdID>::<SerialNo>::INSTR
Use complete VISA resource string.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:DEFine
6.19.6