GUI Reference
R&S
®
ZNA
360
User Manual 1178.6462.02 ─ 12
The "Limit Fail Trace Color" and the appearance of the limit fail symbols are defined in
the
Define User Color Scheme Dialog
("Display" > "Config" > "Define User Color...").
Note:
●
Limit check and display of limit lines are independent of each other:
–
The limit lines can be displayed, no matter if the limit check is enabled.
–
If "Limit Check" is enabled, the limits are checked, no matter if the limit lines
are displayed.
–
The limit check can even be enabled, if no limit lines are defined. In this case,
the info field displays "No limit defined!" and the limit check always passes.
●
Limit lines are defined for a particular trace format. However, the limit check is per-
formed irrespective of the current trace format. The info field indicates the correct
"PASS"/"FAIL" state and limit violations are visualized on the trace (if any).
Figure 6-11: Limit line violations in complex trace formats
Remote command:
CALCulate<Chn>:LIMit:STATe
CALCulate<Chn>:LIMit:LOWer:STATe
CALCulate<Chn>:LIMit:UPPer:STATe
CALCulate<Chn>:LIMit:FAIL?
Lines Softtool