![Rohde & Schwarz 1332.4500.22 User Manual Download Page 1087](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_14777821087.webp)
Command Reference
R&S
®
ZNA
1087
User Manual 1178.6462.02 ─ 12
If the file name is prefixed by "SD:", the file is saved to the SD
card inserted at the calibration unit (if available/accessible).
<TestPort1>
Number of first test port.
<CalUnitPort1>
Number of the calibration unit port to whom the first test port
(<TestPort1>) is assigned.
<TestPort2>
Number of second test port...
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs:ADD
<Characterization>,
<TestPort1>, <CalUnitPort1>, <TestPort2>...
Extends or modifies an
existing
characterization of the active calibration unit (
COMMunicate:RDEVice:AKAL:ADDRess
).
This functionality is not available at the GUI.
Setting parameters:
<Characterization>
Location of an existing characterization (e.g. created using
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs
–
A VNA cal kit file name
*.calkit
without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
–
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1>
Number of first test port.
<CalUnitPort1>
Number of the calibration unit port, to whom test port <Test-
Port1> is assigned.
<TestPort2>
Number of the second test port ...
SCPI Command Reference