Command Reference
R&S
®
ZNA
1331
User Manual 1178.6462.02 ─ 12
[SENSe<Ch>:]FREQuency:MDELay:ACQuire
Starts a calibration sweep for the mixer delay measurement. As a prerequisite, the cali-
brated channel must have at least one mixer delay trace.
Suffix:
<Ch>
.
Channel number
Example:
See
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]FREQuency:MDELay:APERture
<Aperture value>
Defines the frequency difference between the upper and lower tone (aperture).
Suffix:
<Ch>
.
Channel number
Parameters:
<Aperture value>
Aperture
Range:
0 Hz to 50 MHz
*RST:
0 Hz
Default unit: Hz
Example:
*RST; SENS1:FREQ:STAR 1 GHz; STOP 2 GHz
Select a stimulus range between 1 GHz and 2 GHz.
FREQ:MDEL:APER 1 MHz
Select a 1 MHz aperture.
FREQ:MDEL:UTON PORT, 3
Select NWA port 3 as a source port for the upper tone (no exter-
nal generator).
FREQ:MDEL:COMB ON
Use the internal combiner to generate the two-tone signal at port
1 (for instruments equipped with the necessary options).
...
(Perform mixer delay measurement, evaluate results.)
FREQ:MDELay:CONV OFF
Disable mixer delay measurement, switch back to normal (non
frequency-converting) mode.
Manual operation:
See
[SENSe<Ch>:]FREQuency:MDELay:CDELay
<const delay value for ref mixer>
Defines a constant mixer delay value, to be used as a reference for a mixer delay mea-
surement calibration. The setting takes effect after constant mixer delay is enabled;
see example below.
Suffix:
<Ch>
.
Channel number
SCPI Command Reference