Command Reference
R&S
®
ZNA
1255
User Manual 1178.6462.02 ─ 12
It is recommended to use the more general command
.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the fast
calibration mode applies to all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the fast cali-
bration mode is either enabled or disabled for all sources.
Parameters:
<arg0>
Enables (ON) or disables (OFF) fast source power calibration.
*RST:
OFF
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>[:STATe]
<Boolean>
Enables or disables the source power calibration for channel
<Ch>
and for an external
generator number
<Gen>
. The command is disabled unless a source power calibration
for the external generator has been performed (
). To enable or disable a source power calibration for an
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number. This suffix is ignored; the generator is
selected via
<Gen>
.
<Gen>
Generator number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the source power calibration for
port number
<PhyPt>
.
*RST:
OFF
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
SOURce<Ch>:POWer:CORRection:IMODulation:METHod
<FastOrNot>
Enables or disables the fast interpolation method for intermodulation calibration.
Suffix:
<Ch>
.
Channel number
Parameters:
<FastOrNot>
ON (1)
Fast method enabled
OFF (0)
Fast method disabled
*RST:
ON (1)