![Rohde & Schwarz 1332.4500.22 User Manual Download Page 576](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782576.webp)
GUI Reference
R&S
®
ZNA
576
User Manual 1178.6462.02 ─ 12
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:DUT[:STATe]
CALCulate:FMODel:SFD<Ph_pt>:DUT[:STATe]
6.14.5.4
Save Fixture Model
This section allows you to run the selected deembedding tool to generate the Touch-
stone files for fixture deembedding (and the Touchstone file of the de-embedded DUT).
Figure 6-33: Save Fixture Model section
left = single-ended
right = balanced
Run <Fixture Modeling Tool>
Runs the selected fixure deembedding tool.
This button is only active, if both
●
the test coupon measurement finished successfully or the test coupon data were
successfully loaded from file (see
Chapter 6.14.5.2, "Measure Coupon"
●
the measurement of DUT + test fixture finished successfully (see
The resulting Touchstone files (S parameters of the lead-ins, the lead-outs, and the de-
embedded DUT) are written to C:\Users\Public\Documents\Rohde-Schwarz\ZNA
\Embedding. If result files with the same names already exist, you will be asked
whether you want to rename or overwrite them.
Offset Embed Softtool