GUI Reference
R&S
®
ZNA
453
User Manual 1178.6462.02 ─ 12
Messages in the dialog
An information message (or error message) is displayed if one of the following hap-
pens:
●
One of the selected calibration kits is described by ideal kit parameters or typical
values.
●
One of the selected calibration kits does not contain all standards that are required
for the previously selected calibration type.
●
Different connector types are defined at the ports but the selected calibration type
requires uniform connectors.
●
A cal kit standard, cal unit or power meter does not cover the entire calibrated fre-
quency range.
Calibration Type Tab
The "Calibration Type" tab allows you to configure the calibration setup (i.e. a set of
calibration definitions) for the channel selected in the title area of the calibration setup
dialog.
Add
Use "Add" to add a calibration definition to the selected channel. The default calibration
type depends on the channel's measurement mode.
Delete
Select a row in the
table and "Delete" the corresponding calibration defi-
nition.
Cal. Settings
The "Cal. Settings" table displays the calibration definitions of the current channel.
Currently, for each channel either manual calibrations or automatic calibrations are
allowed. When you have configured more than one manual calibration, the CalU-spe-
cific columns disappear:
Port <i> ← Cal. Settings
Tick the checkboxes in the "Port <i>" columns to select the ports to be calibrated in the
corresponding calibration.
Cal Softtool