![Rohde & Schwarz 1332.4500.22 User Manual Download Page 1172](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_14777821172.webp)
Command Reference
R&S
®
ZNA
1172
User Manual 1178.6462.02 ─ 12
<ImaginaryPart>
Imaginary part of the port impedance. This part may be omitted
to define a real impedance.
Range:
- 10 MΩ to 10 MΩ
*RST:
Default reference impedance of the connector fam-
ily assigned to the port (real impedance, e.g. 50 Ω).
Default unit: Ohm
Example:
PORT2:ZREF 52, 2
Specify a complex reference impedance of 52 Ω + j * 2 Ω for the
(physical) port no. 2.
CALC:TRAN:IMP:RNOR PWAV
Select renormalization of port impedances according to the
power waves theory.
Manual operation:
See
"Single Ended Mode / Common Mode / Differential Mode"
8.3.14.20
[SENSe:]POWer...
The
[SENSe:]POWer...
commands configure the optional receiver step attenuators
Chapter 5.7.21, "Receiver Step Attenuators"
on page 226) and the wideband /
narrowband IF gain of the R&S
ZNA.
See also
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
[SENSe<Ch>:]POWer:AGCMode<PhyPt>:MEASure
....................................................... 1172
[SENSe<Ch>:]POWer:IFGain<PhyPt>:MEASure
.............................................................1172
................................................................................1173
................................................................................ 1174
[SENSe<Ch>:]POWer:GAINcontrol:ALL
..........................................................................1174
[SENSe<Ch>:]POWer:GAINcontrol:GLOBal
....................................................................1175
[SENSe<Ch>:]POWer:AGCMode<PhyPt>:MEASure
<Mode>
[SENSe<Ch>:]POWer:IFGain<PhyPt>:MEASure
<AGCMode>
These two equivalent commands activate manual gain control and select the IF gain in
the measurement channels (b-waves). The IF gain in the reference channel (a-waves)
is set automatically if not otherwise defined using
.
This applies to the wideband signal path only.
Note:
These commands are replaced by
and
[SENSe<Ch>:]POWer:GAINcontrol:GLOBal
. They are supported for compatibility
with FW versions < 1.5.
Suffix:
<Ch>
.
Channel number
SCPI Command Reference