Command Reference
R&S
®
ZNA
1136
User Manual 1178.6462.02 ─ 12
Example:
Let's perform a full 3-port calibration with a two-port calibration
unit, using factory characterization and the port assignments
given in the table below.
Prerequisite: the adequate calibration unit was selected before
using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
SENSe1:CORRection:COLLect:AUTO:ASSignment:
DELete:ALL
Deletes all available port assignments.
SENSe1:CORRection:COLLect:AUTO:CONFigure
'FNPort',''
Sets the automatic calibration to "Full n-port" with factory char-
acterization.
SENSe1:CORRection:COLLect:AUTO:ASSignment1:
DEFine 2,1,3,2
Creates port assignment 1 explicitely (no auto-detection).
SENSe1:CORRection:COLLect:AUTO:ASSignment2:
DEFine:TPORt 3,4
Creates port assignment 2 implicitely (auto-detection).
Before starting the calibration sweep for port assignment 1,
ensure test port 2 is connected to cal unit port 1 and test port 3
to cal unit port 2
SENSe1:CORRection:COLLect:AUTO:ASSignment1:
ACQUire
Perfoms the calibration sweep for port assignment 1
Before starting the calibration sweep for port assignment 2,
ensure test ports 3 and 4 are connected to the cal unit (in any
order)
SENSe1:CORRection:COLLect:AUTO:ASSignment2:
ACQUire
Perfoms the calibration sweep for port assignment 2; auto-
detects the existing port-connections at runtime
SENSe1:CORRection:COLLect:AUTO:SAVE
Checks whether the acquired calibration data are sufficient to
calculate the system error correction. If yes, applies them to the
selected channel.
See
Chapter 9.2.5.2, "MultiCal (with Calibration Unit)"
on page 1395 for a MultiCal example.
Manual operation:
See
Test Port
Port Assignment 1
Port Assignment 2
2
Cal Unit Port 1
-
3
Cal Unit Port 2
auto-detected
4
-
auto-detected
SCPI Command Reference