Command Reference
R&S
®
ZNA
1111
User Manual 1178.6462.02 ─ 12
the position until position 20 has been recorded. Subsequent calls will start over at
position 1, overwriting the previously loaded data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<CalGroupFile>
String parameter to specify the name of the loaded cal group
file.
<Standard>
THRough | OPEN | SHORt | OSHort | OSHort1 | OSHort2 |
OSHort3 | MATCh | NET | ATT | REFL | SLIDe | ISOLation |
LINE | LINe1 | LINe2 | LINe3 | UTHRough | POWer | MIXer
Standard types: Through, Open, Short, Match, Symmetric Net-
work (NET), Attenuation (ATT), Reflect, Sliding Match (SLIDe),
Line1 (LINE1 and LINE are synonymous), Line2 and Line3 (esp.
for TRL calibration), Offset Short 1 to 3 (OSHort), Unknown
Through, power meter (for SMARTerCal), reference mixer
ISOLation is not a physical standard: to measure the isolation
(supported for transmission normalization and TOSM only!) it is
recommended to suitably terminate the related test ports (e.g.
with 50 Ω loads).
<TestPort1>
Test port number. For a transmission standard (through, line,
attenuation, symmetric network) or an adapter used as a
"through" the input and output port numbers must be specified.
For reflection standards, only one port number is required.
Setting parameters:
<SecondPortOrAdapter>
For a transmission standard or an adapter used as a "through"
this parameter specifies the second test port.
For reflection measurements with an adapter connected
between port and standard, set it to ON.
Otherwise set it to OFF or simply omit it.
Example:
Suppose that the cal pool contains a file
'Calgroup3.cal'
with a valid through calibration for the active channel no. 1,
which you want to include in a new TOSM calibration for ports 1
and 2.
MMEM:CDIR DEF; CDIR 'Calibration\Data'
Go to the cal group directory.
:SENSe1:CORRection:COLLect:METHod:DEFine 'New
Cal', TOSM, 1, 2
Define a new TOSM calibration.
SENSe1:CORRection:COLLect:LOAD:SELected 'Cal
Group 3.cal', THROUGH, 1, 2
Load the through data from the cal pool file into the new calibra-
tion.
SENSe1:CORRection:COLLect:ACQuire:SELected
OPEN, 1
Proceed with the new calibration measurements.
Manual operation:
See
SCPI Command Reference