![Rohde & Schwarz 1332.4500.22 User Manual Download Page 1356](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_14777821356.webp)
Command Reference
R&S
®
ZNA
1356
User Manual 1178.6462.02 ─ 12
Example:
CORR:CKIT:INST 'C:\Users\Public\Documents
\Rohde-Schwarz\ZNA\Calibration\Kits\ZCAN.ck'
Load the previously created R&S ZVR cal kit file
ZCAN.ck
from
the default cal kit directory.
MMEM:STOR:CKIT 'ZCAN', 'C:
\Users\Public\Documents
\Rohde-Schwarz\ZNA\Calibration\Kits\ZCAN.calkit'
Store the imported cal kit data to a VNA cal kit file
ZCAN.calkit
(assuming that the cal kit name stored in
ZCAN.ck
reads "ZCAN").
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]
<Standard>[, <Dispersion>,
<Delay>]
Starts a calibration measurement in order to acquire measurement data for the
selected standards. The standards are reflection or transmission standards and must
be connected to port 1 or 2 of the analyzer.
Tip:
[SENSe<Ch>:]CORRection:COLLect[:
to obtain measurement data at arbitrary analyzer ports.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
Setting parameters:
<Standard>
THRough | OPEN1 | OPEN2 | OPEN12 | SHORt1 | SHORt2 |
SHORt12 | MATCh1 | MATCh2 | MATCh12 | NET | ATT |
IMATch12 | REFL1 | REFL2 | SLIDe1 | SLIDe2 | SLIDe12 |
LINe1 | LINe2 | LINe3 | M1O2 | O1M2 | OSHort1 | OSHort11 |
OSHort12 | OSHort13 | OSHort2 | OSHort21 | OSHort22 |
OSHort23 | M1S2 | S1M2 | UTHRough
Standard types: Through (between port 1 and 2), Open, Short,
Match (MATCh12 and IMATch12 are synonymous), Symmetric
Network (NET), Attenuation (ATT), Reflect, Sliding Match
(SLIDe), Line1 (LINe1 and LINE are synonymous), Line2 and
Line3 (esp. for TRL calibration), Match/Open (M1O2, O1M2),
Match/Short (M1S2, S1M2), Offset Short (OSHort), Unknown
Through (UTHRough).
The numbers in the parameter names denote the analyzer ports.
Two numbers 12 mean that two separate calibrations are per-
formed at ports 1 and 2. For Offset Short standards, the first
number denotes the port (1 or 2), the second number denotes
the number of the standard (1 to 3).
*RST:
ON
<Dispersion>
Optional status parameter for UTHRough standard:
OFF - unknown through standard is non-dispersive.
ON - unknown through standard is dispersive.
*RST:
OFF
R&S ZVR/ZVABT Compatible Commands