GUI Reference
R&S
®
ZNA
470
User Manual 1178.6462.02 ─ 12
button to open the
, which allows you to
add calibration definitions or modify them. Existing definitions can be selectively
.
Use the checkboxes in the "Enable" column to enable/disable the related calibration
definition. Use the ellipsis button in the "Channels" column to assign the related cali-
bration definition (also) to other channels.
Add + Modify
that allows you to add calibration definitions and
modify existing ones.
Delete
Deletes the calibration definition that is selected in the
"Calibration definition table"
Port table
The table in the lower half of the dialog displays the port configuration of the item
selected in the row in the
●
For an automatic calibration (left), it displays the port assignment.
●
For a manual calibration (right), it displays the ports with their connector, gender
and cal kit.
For a given port, these values are shared among the related manual channel cali-
bration definitions.
The buttons "Same Connector all Ports", "Same Gender all Ports" and "Import Cal
Kit..." provide the functionality known from the
Start Cal
Saves the calibration definitions to the current setup. Then enters the
to acquire data for all involved channels, ports and standards.
Note:
In the current implementation, "Start Cal" is disabled if both manual and auto-
matic calibration definitions are enabled in the
.
Cal Softtool