Getting Started
R&S
®
ZNA
59
User Manual 1178.6462.02 ─ 12
6. Select "OK" to apply your settings and return to the calibration setup dialog "Cali-
bration Setting" wizard.
7. Select the test port connector type and gender (here: N 50 Ω, female, correspond-
ing to a male Through standard), and the calibration kit (here: R&S
ZV-Z121).
8. Tap "Start Cal".
9. The calibration dock widget indicates the standard measurements that make up a
"Trans Norm" calibration.
Select "Through (mm)" to initiate the measurement of the connected Through stan-
dard. Measuring the isolation between ports 1 and 2 is optional. Skip it for now.
Performing Measurements