Command Reference
R&S
®
ZNA
1176
User Manual 1178.6462.02 ─ 12
[SENSe<Ch>:]PULSe:ATIMe
<Acquisition time>
Defines the duration of the data acquisition for pulse measurements in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Acquisition time>
Data acquisition time
Default unit: s
Manual operation:
See
[SENSe<Ch>:]PULSe:COUPled[:STATe]
<Boolean>
Enforces identical pulse generator settings for all ports.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
*RST:
OFF (0)
Options:
R&S
ZNA<Freq>-B4<Pt> or R&S
ZNA-B91
Manual operation:
See
[SENSe<Ch>:]PULSe:DUTYcycle
<Percentage>
Defines the duty cycle of the generated pulses.
This setting only takes effect, if:
●
Internal Pulse Control is enabled (
[SENSe<Ch>:]PULSe:INTernal[:STATe]
●
Fixed Duty Cycle is enabled (
[SENSe<Ch>:]PULSe:FXDCycle[:STATe] ON
Note
that currently the exact value of the duty cycle can not be guaranteed. However,
you can be sure that the specified value is not exceeded.
Suffix:
<Ch>
.
Channel number
Parameters:
<Pulse Duty Cycle>
A value between 0 and 100.
Default unit: percent
Options:
R&S
ZNA<Freq>-B4<Pt> or R&S
ZNA-B91
Manual operation:
See
[SENSe<Ch>:]PULSe:FXDCycle[:STATe]
<Boolean>
Defines whether the analyzer firmware shall enforce a fixed duty cycle for all pulse
generators.
SCPI Command Reference