![Rohde & Schwarz 1332.4500.22 User Manual Download Page 1425](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_14777821425.webp)
Annexes
R&S
®
ZNA
1425
User Manual 1178.6462.02 ─ 12
1. Run the R&S
ZNA and configure it as required for the intended showroom opera-
tion.
2. Save the configuration as described in
Chapter 6.15.1, "Recall Sets Tab"
on page 593. If necessary, move the recall set file (
*.znxml|*.znx
) to the appro-
priate location.
3. Under the registry key
HKEY_LOCAL_MACHINE/SOFTWARE/Rohde-Schwarz/ZNA
, create a new subkey
ShowroomMode
.
4. Within the
ShowroomMode
-subkey, create the string value
ShowroomSetup
.
5. Modify the value
ShowroomSetup
: set its value data to the full path of the recall
set. Use forward slashes as path separators, e.g. "C:/Documents and Settings/All
Users/Rohde-Schwarz/ZNA/RecallSets/ShowroomSet1.znxml".
After the analyzer is restarted or "Preset", the specified recall set is loaded.
11.6
ENA Emulation Commands
The following table lists the commands implemented/overwritten for the ENA remote
language.
ENA Emulation Command
R&S
ZNA Default Parser Command
CALCulate<Ch>:FSIMulator:SENDed:DEEM-
bed:PORT<Pt>[:TYPE] {USER|NONE}
CALCulate<Ch>:TRANsform:VNETworks:
SENDed:DEEMbedding<PhyPt>:TNDefinition
CALCulate<Ch>:FSIMulator:SENDed:DEEM-
bed:PORT<Pt>:USER:FILename <string>
CALCulate<Ch>:TRANsform:VNETworks:
SENDed:DEEMbedding<PhyPt>:PARameters:
DATA
CALCulate<Ch>:FSIMulator:SENDed:DEEM-
bed:STATe {ON|OFF|1|0}
n.a. (enabled channel-by-channel, port-by-port)
CALCulate{1..4}:FSIMulator:SENDed:PMCir-
cuit:STAT {ON|OFF|1|0}
CALCulate<Ch>:TRANsform:VNETworks:
SENDed:EMBedding<PhyPt>[:STATe]
CALCulate{1..4}:FSIMulator:SENDed:PMCir-
cuit:PORT<Pt>[:TYPE] {NONE|PCSC|PCSL|PLPC|
PLSC|PLSL|SCPC|SCPL|SLPC|SLPL|USER}
CALCulate<Ch>:TRANsform:VNETworks:
SENDed:EMBedding<PhyPt>:TNDefinition
CALCulate{1..4}:FSIMulator:SENDed:PMCir-
cuit:PORT<i>:PARameter:C <numeric>
CALCulate<Ch>:TRANsform:VNETworks:
SENDed:EMBedding<PhyPt>:PARameters:
C<Cmp>
CALCulate{1..4}:FSIMulator:SENDed:PMCir-
cuit:PORT<i>:PARameter:G <numeric>
CALCulate<Ch>:TRANsform:VNETworks:
SENDed:EMBedding<PhyPt>:PARameters:
G<Cmp>
CALCulate{1..4}:FSIMulator:SENDed:PMCir-
cuit:PORT<i>:PARameter:L <numeric>
CALCulate<Ch>:TRANsform:VNETworks:
SENDed:EMBedding<PhyPt>:PARameters:
L<Cmp>
ENA Emulation Commands