![Rohde & Schwarz 1332.4500.22 User Manual Download Page 787](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782787.webp)
Command Reference
R&S
®
ZNA
787
User Manual 1178.6462.02 ─ 12
This tolerance will only be used if
CALCulate<Chn>:EYE:MASK:FAIL:CONDition
is set to SAMPles.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<ViolationTolerance>
Options:
R&S
ZNA-K20
Manual operation:
See
CALCulate<Chn>:EYE:MEASurement:DATA?
Returns the measurement results of the related eye diagram (see
The return value is of type string and returns the eye measurement results in csv for-
mat with decimal separator "." and field separator "," (see also
).
Note
that the full set of measurement results is only available for NRZ modulated gen-
erator signals (see
CALCulate<Chn>:EYE:INPut:MODulation
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI Command Reference