![Rohde & Schwarz 1332.4500.22 User Manual Download Page 782](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782782.webp)
Command Reference
R&S
®
ZNA
782
User Manual 1178.6462.02 ─ 12
Usage:
Query only
Options:
R&S
ZNA-K20
Manual operation:
See
CALCulate<Chn>:EYE:MASK:FAIL?
Returns
'Pass'
or
'Fail'
to indicate the result of the limit check in the related eye
diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram, if the eye test is not enabled (
ON), or if the mask is empty (see
late<Chn>:EYE:MASK:SHAPe:BOTTom|TOP|POLYgon:STATe
.
Tip
●
to perform a composite (global) limit check
●
The result is automatically recalculated whenever a relevant setting is changed, i.e.
a subsequent query will return the updated mask test result
Suffix:
<Chn>
.
Channel number used to identify the active trace
Return values:
<TestResults>
Usage:
Query only
Options:
R&S
ZNA-K20
Manual operation:
See
CALCulate<Chn>:EYE:MASK:FAIL:BEEP
<Boolean>
Defines whether the R&S
ZNA should make an audible beep on mask failures in the
related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
*RST:
OFF
Options:
R&S
ZNA-K20
Manual operation:
See
CALCulate<Chn>:EYE:MASK:FAIL:CONDition
<FailCondition>
Defines whether the fail condition for the eye mask test of the related eye diagram is
specified in absolute or relative terms.
SCPI Command Reference