GUI Reference
R&S
®
ZNA
404
User Manual 1178.6462.02 ─ 12
6.8.2.1
Controls on the Power Tab
For power sweeps, the
"Stimulus" softtool.
Some of the buttons on the "S-Params" tab open related dialogs.
●
Chapter 6.8.2.2, "Power Config Dialog"
●
Chapter 6.8.2.3, "ALC Config Dialog"
Power
Determines the output power at the test ports for the sweep types "Lin Freq", "Log
Freq", "CW Mode", and "Time". Also determines the default output power for "Segmen-
ted" sweeps, where the output power can be set per segment.
The setting has no effect for
sweeps, where the source power is varied over a
continuous range.
Remote command:
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]
Source Step Att.
This section is only available if the R&S
ZNA is equipped with mechanical source step
attenuators (see
Chapter 5.7.19, "Source Step Attenuators"
Attenuation can be set from 0 to 70 dB, in steps of 10 dB.
Remote command:
SOURce<Ch>:POWer<PhyPt>:ATTenuation
Stimulus Softtool