Concepts and Features
R&S
®
ZNA
228
User Manual 1178.6462.02 ─ 12
Source
Ref
Meas
Src
Mon
Out
Out
In
Src
Step
Att
Rec
Step
Att
Port y = 1, 3
In
Out
Ref
Source Meas
Out
Out
In
In
Port y
Src Mon Out
Figure 5-28: B16 jumpering for direct source monitoring
●
Options R&S
ZNAxx-B161/163 are only available with hardware option
R&S
ZNAxx-B16 (see
Chapter 5.7.17, "Direct Generator/Receiver Access"
R&S
ZNAxx-B161 requires R&S
ZNAxx-B21
–
R&S
ZNAxx-B163 requires R&S
ZNAxx-B21
and
R&S
ZNAxx-B23
(see
Chapter 5.7.19, "Source Step Attenuators"
●
To retrofit a direct source monitor access, the R&S
ZNA must be sent to
Rohde
&
Schwarz service.
5.7.25
Internal Combiner P1–P3
Option R&S
ZNAxx-B213
On a four-port R&S
ZNAxx with hardware option B213, the (independent) source sig-
nals generated at physical ports 1 and 3 can be internally combined and routed to
port
1.
No external combiner or additional cabling is needed. The performance loss of
is avoided.
An internal combiner is particularly useful for
and
Embedded LO Mixer Group Delay Measurements
, that require a two-tone signal.
It is possible to switch the combiner on or off between channels.