Command Reference
R&S
®
ZNA
1125
User Manual 1178.6462.02 ─ 12
Example:
See
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation:
See
"Loss at DC / Loss at Freq / Freq for Loss"
[SENSe<Ch>:]CORRection:NFIGure[:STATe]
<Boolean>
Disables or enables the Noise Figure Calibration for the related channel.
Suffix:
<Ch>
.
Parameters:
<Boolean>
OFF - Disable calibration
ON - (re-)enable calibration
*RST:
OFF (calibration disabled, no noise figure calibra-
tion available after a reset)
Options:
R&S
ZNA-K30
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>[:STATe]
<Boolean>
Resets the offset parameters for all test ports to zero and the reference frequency to 1
GHz or queries whether any of the offset parameters are different from zero.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer. This numeric suffix is ignored; the
command affects the parameters of all ports.
Parameters:
<Boolean>
The parameter function depends on whether the command is
used as a setting command or as a query: For setting command:
ON - no effect
OFF - resets all length offsets to zero and the reference fre-
quency to 1 GHz
For query:
1 - at least one length offset parameter is different from its
default value
0 - all length offsets are zero / set to default
*RST:
OFF
Example:
*RST; :CORR:OFFS?
Reset the instrument and query whether the length offset param-
eters have been reset as well. The response is 0.
Manual operation:
See
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe]
<Boolean>
Toggles length/loss/fixture compensation for physical port <PhyPt> ON/OFF.
SCPI Command Reference