![Rohde & Schwarz 1332.4500.22 User Manual Download Page 578](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782578.webp)
GUI Reference
R&S
®
ZNA
578
User Manual 1178.6462.02 ─ 12
Test Coupons > Insertion Loss
Tells the ISD tool about the linearity of the 2xThru test coupon:
●
"Linear": linear insertion loss
●
"Non-Linear" (default): non-linear insertion loss
●
"Resonant": the 2x thru test coupon will be split and used directly for deembedding
This option may be more accurate when the fixture and 2x Thru have the same
impedance at every location.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:ATTenuation:BEHavior
Lead Ins
Defines how the ISD tool shall interpret the lead-in data.
Scaling for Lead In Atten ← Lead Ins
Scales the test coupon’s attenuation.
Default is 1.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:SCALe:ATTenuation
Offset Embed Softtool