GUI Reference
R&S
®
ZNA
547
User Manual 1178.6462.02 ─ 12
PuMo In
is enabled, this setting allows you to select
between positive or negative polarity of the "PuMo In" signal.
Remote command:
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:INPut:INVerted[:STATe]
Trig. Out
Allows you to enable/disable pulse modulator control via the respective "Trig. Out".
Note
that a Trigger Out that has been previously configured to indicate sweeps, seg-
ments or measurement points for this channel, cannot be used for pulse generation.
The corresponding table row is disabled in this case.
To change the trigger configuration, use the "Trigger Out Manager..." button to open
the
.
Remote command:
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
PGENerator | NONE
TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe
POSitive | NEGative
Measurement Settings
Sets up the "Pulse Profile" measurement for this channel (see
Acq. Delay/Fixed
Defines the acquisition delay relative to the master pulse. Can only be set in synchron-
ized measurement mode.
Remote command:
Acq. Time/Fixed
Defines the duration of the data acquisition.
Remote command:
Bandwidth/Fixed
Defines the resolution bandwidth.
Remote command:
[SENSe<Ch>:]BWIDth[:RESolution]
Sync Meas to Pulse Gen.
Allows you to switch between synchronous and asynchronous measurement.
Remote command:
[SENSe<Ch>:]PULSe:SYNCron[:STATe]
Autoset
Adjusts "Acq. Delay" and "Acq. Time" according to the configured pulses.
Mode Softtool