![Rohde & Schwarz 1332.4500.22 User Manual Download Page 1055](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_14777821055.webp)
Command Reference
R&S
®
ZNA
1055
User Manual 1178.6462.02 ─ 12
Setting parameters:
<Type>
The name of the new converter type. Must be unique among the
existing ones (see
Example:
SENSe:CONVerter:DEFinition:DEFine 'My converter
type'
creates a new converter type named "My converter type".
SENSe:CONVerter:DEFinition:CATalog?
returns something like
'ZVA-Z110',...,'My converter
type'
Usage:
Setting only
Options:
R&S
ZNA-K8
Manual operation:
See
[SENSe:]CONVerter:DEFinition:DELete
<Type>
Deletes the related converter type.
Only user-defined converter types can be deleted.
Setting parameters:
<Type>
Name of the converter type to be deleted.
Usage:
Setting only
Options:
R&S
ZNA-K8
Manual operation:
See
[SENSe:]CONVerter:DEFinition:FREQuency
<Type>, <StartFreq>, <StopFreq>
[SENSe:]CONVerter:DEFinition:FREQuency?
<Type>
Sets/queries the start and stop frequency of (the converter port of) the related con-
verter type.
Parameters:
<StartFreq>
Default unit: Hz
<StopFreq>
Default unit: Hz
Parameters for setting and query:
<Type>
Name of the converter type, whose start and stop frequencies
are set/queried.
Options:
R&S
ZNA-K8
Manual operation:
See
"Converter Port: Start Freq./Stop Freq."
SCPI Command Reference