GUI Reference
R&S
®
ZNA
265
User Manual 1178.6462.02 ─ 12
b2 2nd Harmonic Src Port 1
Power of the outgoing wave at port 2, measured at 2 times the frequency of the wave
generated at port 1.
Remote command:
CALCulate<Ch>:PARameter:MEASure
'<TraceName>', 'h2b2(p1)'
CALCulate<Ch>:PARameter:SDEFine
'<TraceName>', 'h2b2(p1)'
...
b2 3rd Harmonic Src Port 1
Power of the outgoing wave at port 2, measured at 2 times the frequency of the wave
generated at port 1
Remote command:
CALCulate<Ch>:PARameter:MEASure
'<TraceName>', 'h3b2(p1)'
CALCulate<Ch>:PARameter:SDEFine
'<TraceName>', 'h3b2(p1)'
...
Wave Correction
Determines whether wave quantities and ratios are system error corrected in the active
channel.
If checked (default), the "best available" calibration is applied, with the following prece-
dence:
●
Scalar power calibration
●
User system error correction
●
Factory system error correction
Note:
●
Only a single calibration is applied.
●
If "Wave Correction" is turned off, no calibration is applied for these measurement
quantities at all.
Remote command:
[SENSe<Ch>:]CORRection:EWAVe[:STATe]
6.2.5.2
More Harmonics Dialog
The controls in the "More Harmonics" dialog allow you to select and configure arbitrary
harmonics measurements in the current channel.
Access:
Trace – [Meas] > "S-Params" > "Harmonics" > "More Harmonics..."
Meas Softtool