![Rohde & Schwarz 1332.4500.22 User Manual Download Page 427](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782427.webp)
GUI Reference
R&S
®
ZNA
427
User Manual 1178.6462.02 ─ 12
Figure 6-14: Point List
Import.../ Export...
The buttons open standard dialogs to import/export sweep segment settings to/from an
ASCII file (
*.SegList
).
●
"Import..." replaces the current segment list by a sweep segment list loaded from a
*.SegList
file.
●
"Export..." stores the current sweep segments settings to a
*.SegList
file.
Sweep segment files
The analyzer uses a simple ASCII format to export sweep segment data. By default,
the sweep segment file extension is
*.SegList
. The file starts with two comment
lines containing the version and a third line reproducing the header of the segment list.
The following lines contain the entries of all columns of the segment list, including the
"Displayed Columns" that ae configured in the "Define Segments" dialog.
Example:
The segmented sweep range
is described by the following sweep segment file:
Note:
The
*.SegList
file actually contains more columns listing all channel settings
of the individual sweep segments. The headings of the additional columns are
IF Band-
width [Hz]
,
en:IF Selectivity
,
en:IF Sideband
,
Meas Delay [µs]
,
bo:Sweep Time
Auto
,
en:Frq Sweep Mode
.
Remote command:
MMEMory:LOAD:SEGMent
MMEMory:STORe:SEGMent
Sweep Softtool