Command Reference
R&S
®
ZNA
802
User Manual 1178.6462.02 ─ 12
Parameters:
<Boolean>
Manual operation:
See
"Automatic Flt Tm for DUT + Lead Ins"
CALCulate:FMODel:ISD<Ph_pt>:OPERation
<Operation>
Defines the execution mode of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Operation>
FAST | ACCurate
ACCurate: normal execution mode
FAST: reduces the execution time to ~50%
In many cases the FAST mode can be enabled with only little
loss of accuracy.
Manual operation:
See
CALCulate:FMODel:ISD<Ph_pt>:PASSivity
<Boolean>
Defines whether the ISD tool shall enforce passivity and reciprocity for the test cou-
pons and the test fixture.
Corresponds to the
passive
batch mode parameter of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Boolean>
ON (0): enforce passivity and reciprocity
OFF (0): do not enforce passivity and reciprocity
Manual operation:
See
CALCulate:FMODel:ISD<Ph_pt>:PORT:ORDer
<PortOrder>
Defines how the ISD tool shall interpret the DUT + Fixture data (see
):
●
NON: ports 1 to N are on the left and ports N+1 to 2*N are on the right
●
ODD: ports 1, 3, 5, etc. are on the left and ports 2, 4, 6, etc. are on the right.
●
ALL: use this to tell the ISD tool that all ports are on the left (i.e. assumed to be
coupled)
Corresponds to the
port_order
batch mode parameter of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<PortOrder>
ODD | NON | ALL
SCPI Command Reference