User Manual 1178.6462.02 ─ 12
To retrofit the memory extension board, the R&S
ZNA must be sent to Rohde &
Schwarz Service.
5.7.16
LO Out
R&S
ZNA-B8
Hardware option R&S
local oscillator output
port.
This port is particularly useful for driving millimeter wave converters. If you use a power
splitter to connect the LO Out of the R&S
ZNA to the LO In ports of the converters,
then it is possible to measure n converter ports with an n-port R&S
ZNA, without the
need for an external generator.
5.7.17
Direct Generator/Receiver Access
R&S
ZNAxx-B16
Hardware option R&S
ZNAxx-B16 provides 3 pairs of SMA connectors for each test
port of an analyzer R&S
ZNAxx.
See
"Direct Generator and Receiver Access"
Reverse Coupler Configuration
The reverse coupler configuration is obtained by a simple reconnection of the direct
generator/receiver access jumpers.
By inverting the signal routing of RF out and measurement path, the sensitivity of the
measurement receiver is increased by 10 to 20
dB.