![Rohde & Schwarz 1332.4500.22 User Manual Download Page 963](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782963.webp)
Command Reference
R&S
®
ZNA
963
User Manual 1178.6462.02 ─ 12
Parameters:
<Boolean>
0 – Marker <Mk> of trace <TraceName> is displayed
1 – Marker <Mk> of trace <TraceName> is hidden
Setting parameters:
<TraceName>
Name of the related trace.
Manual operation:
See
DISPlay[:WINDow<Wnd>][:STATe]
<Boolean>
Creates or deletes a diagram area, identified by its area number
<Wnd>
.
Suffix:
<Wnd>
.
Number of the diagram area to be created or deleted.
Parameters:
<Boolean>
ON | OFF - creates or deletes diagram area no.
<Wnd>
.
*RST:
-
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
Manual operation:
See
DISPlay[:WINDow<Wnd>]:CATalog?
Returns the numbers and names of all diagrams in the current recall set.
The response is a string containing a comma-separated list of diagram area numbers
and names, see example below. If all diagram areas have been deleted, the response
is an empty string ("").
Suffix:
<Wnd>
.
Number of a diagram. This suffix is ignored; the command
returns a list of all diagrams.
SCPI Command Reference