![Rohde & Schwarz 1332.4500.22 User Manual Download Page 781](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782781.webp)
Command Reference
R&S
®
ZNA
781
User Manual 1178.6462.02 ─ 12
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<VerticalOffset>
Vertical offset relative to the 0 V level, i.e. the Voltage level of
the center.
Default unit: V
Options:
R&S
ZNA-K20
Manual operation:
See
CALCulate<Chn>:EYE:MASK:DATA?
Returns the detailed results of the mask test in the related eye diagram (i.e. the con-
tents of the corresponding result info field).
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram, or if
is
OFF
.
Use
MMEMory:STORe:EYE:MASK:RESults
to save these results to an ASCII file.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
CALCulate1:EYE:MASK:DATA?
returns something like
'Eye Mask
------------------------------
Fail Condition Type Samples
Violation Tolerance 1
Total Number of Samples 10342
Mask 1 (Top) Active
Samples Hits 366
Fail Rate 3.539 %
Test Result Fail
Mask 2 (Bottom) Not Active
Samples Hits ------
Fail Rate ------
Test Result ------
Mask 3 (Octagon) Not Active
Samples Hits ------
Fail Rate ------
Test Result ------
'
SCPI Command Reference