![Rohde & Schwarz 1332.4500.22 User Manual Download Page 896](http://html1.mh-extra.com/html/rohde-and-schwarz/1332-4500-22/1332-4500-22_user-manual_1477782896.webp)
Command Reference
R&S
®
ZNA
896
User Manual 1178.6462.02 ─ 12
Parameters:
<Boolean>
ON - continuous extrapolation enabled
OFF - continuous extrapolation disabled
*RST:
OFF
Example:
See
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:EXTRapolate
Extrapolates the measured trace towards f = 0 and overwrites the current DC value
(
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam
). The command is rele-
vant for low pass time domain transforms.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
See
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam
Usage:
Event
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:LPFRequency
Calculates the harmonic grid for low pass time domain transforms, keeping the stop
frequency and the number of points.
Tip:
CALCulate<Chn>:TRANsform:TIME:LPASs
if you wish to use one of the
other algorithms for calculating the grid.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
See
CALCulate<Chn>:TRANsform:TIME[:TYPE]
Usage:
Event
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor
<REfactor>
Defines the resolution enhancement factor for the time domain transform.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<REfactor>
Resolution enhancement factor.
Range:
1 to 10.
Increment: 0.1
*RST:
1 (no resolution enhancement)
Example:
See
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam
SCPI Command Reference