9-23
115-056332-00
Air supply inlet assembly
(0635/DISS/electronic pressure monitoring)
DISS/electronic pressure
monitoring
9.2.23 Remove Backup Gas Source Inlet Port Assembly
1. Remove the service door and rear cover plate.
2. Disconnect the pipeline connected with the backup gas source inlet port assembly.
3. Unscrew the four (4) screws fixing the metal plate of backup gas source, and remove the
backup gas source assembly.
4. Unscrew the two (2) screws fixing each gas source inlet port assembly to remove one
backup gas source inlet port assembly.
Refer to the following table for backup oxygen/pipe negative pressure inlet assemblies and
cylinder inlet assemblies:
P/N Description
Application
115-034706-01
Backup oxygen supply assembly (0615/umbrella
valve)
Backup oxygen/without
electronic pressure
monitoring
115-030793-01
Backup oxygen inlet assembly
(NIST/0625/umbrella valve)
Backup oxygen\with
electronic pressure
monitoring
115-004519-01 O
2
gas supply inlet assembly (with cylinder)
NIST/with cylinder
115-004520-01 N
2
O gas supply inlet assembly (with cylinder)
NIST/with cylinder
115-004558-01 O
2
gas supply inlet assembly (DISS/with cylinder)
DISS/with cylinder
115-004559-01
N
2
O gas supply inlet assembly (DISS/with
cylinder) DISS/with
cylinder
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...