7-33
9. Select the required calibration device, select [
Start
]. The calibration interface as shown
in the following figure is displayed. During the calibration process, you can select
[
Cancel
] to cancel calibration.
10. The interface as shown in the following figure is displayed after the calibration is
complete.
If the pressure sensor calibration fails, the interface as shown in the following figure is
displayed. Select [
Retry
] to re-calibrate. Select [
OK
] to quit calibration.
If the pressure sensor calibration succeeds, the interface as shown in the following figure
is displayed. Select [
OK
] to quit calibration.