8-20
Message
Alarm
Priority
Cause
Solution
BIS R Over
Range
Low
The monitored value of
BIS R exceeds the
measurable range.
Replace the BIS module.
BIS High
Imped.
Low
The sensor is not properly
attached to the skin of the
patient, resulting in a high
impedance.
1. Check whether the BIS sensor is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Sensor Off Low
The sensor is detached
from the patient.
1. Check whether the BIS sensor is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
1 High Imped.
Low
The impedance of BIS
electrode 1 is too high.
1. Check whether BIS electrode 1 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
1 Lead Off
Low
BIS electrode 1 is
detached from the patient.
1. Check whether BIS electrode 1 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
2 High Imped.
Low
The impedance of BIS
electrode 2 is too high.
1. Check whether BIS electrode 2 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
2 Lead Off
Low
BIS electrode 2 is
detached from the patient.
1. Check whether BIS electrode 2 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
3 High Imped.
Low
The impedance of BIS
electrode 3 is too high.
1. Check whether BIS electrode 3 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
3 Lead Off
Low
BIS electrode 3 is
detached from the patient.
1. Check whether BIS electrode 3 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
4 High Imped.
Low
The impedance of BIS
electrode 4 is too high.
1. Check whether BIS electrode 4 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
4 Lead Off
Low
BIS electrode 4 is
detached from the patient.
1. Check whether BIS electrode 4 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
G High Imped.
Low
The impedance of BIS
electrode G is too high.
1. Check whether BIS electrode G is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
G Lead Off
Low
BIS electrode G is
detached from the patient.
1. Check whether BIS electrode G is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
Low
The impedance of BIS
1. Check whether BIS electrode C is
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...