8-86
8. Turn on the pipeline N2O source and record the reading on the test pressure gauge. If
the difference between the readings of the test pressure gauge and N2O pipeline pressure
gauge exceeds 0.1 MPa (1 bar), the N2O pipeline pressure gauge is damaged. Handle
this problem according to the fault identification table.
9. Reconnect the pipeline to the pressure gauge. Reconnect pipeline 40.
III. Air supply inlet:
1. Turn off the pipeline gas source. Disconnect pipeline 50. Pull out the pipeline end that
connects to the system switch block.
2. Connect the 1 MPa test pressure gauge to the system switch by using the 3106-04-06
adapter connector and Y-shape connector 3140-06-00.
3. Turn on the pipeline air source. If the reading on the test pressure gauge is not within
0.2-0.25 MPa (2.0-2.5 bar), adjust the regulator to ensure that the reading on the test
pressure gauge is 0.2 MPa (2.0 bar). Record the reading on the air pipeline pressure
gauge.
4. Turn off the pipeline air source and turn on the air flow regulator to release the
remaining pressure.
5. Reconnect
pipeline
50.
6. Disconnect PU tube 41 that connects the air supply inlet and the air pipeline pressure
gauge. Pull out the pipeline end that connects to the air supply inlet.
7. Connect the 1 Mpa test pressure gauge to the air supply outlet by using the 3106-04-06
adapter connector.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...