8-2
8.1.4 Cleanup of the Maintenance Area
Clean up the maintenance area after each maintenance.
8.2 Technical Alarms
A technical alarm, as apposed to a parameter alarm, is an alarm condition that exists no matter
whether a patient is connected to the machine. Technical alarms include:
Startup alarm messages
CPU board runtime alarms
Power board runtime alarms
Runtime alarms of the flow sensor board
VCM runtime alarms
AG module runtime alarms
Before troubleshooting the anesthesia machine, check for technical alarm messages. If an alarm
message is presented, eliminate the alarm first.
The following sections detail how to troubleshoot technical alarms related to the modules
mentioned above.
For detailed information on possible causes and actions for other alarm and prompt messages, see
the Operator’s Manual.
8.2.1 Startup Alarm Messages
Message
Alarm
Priority
Cause
Solution
Bundle Version
Error
Bundle Version:
Time out
High
1. The software versions are
not compatible with each
other.
2. "Time out" is displayed
when the communication
between the corresponding
software and the system
software fails.
1. Re-upgrade the compatible software
version.
Flowmeter Self
Test Error
Flowmeter Self
Test: Time out
High
1. A leak exists on the
O2/N2O/AIR airway.
2. An error occurs during
flowmeter zeroing.
1. Restart the machine for a self test
again.
2. If the problem persists, replace the
electronic flowmeter assembly.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...