3-6
13. Rotate the casters for 90°. Then push this device gently to roll it down the slope.
Remove the packing box on the equipment. Keep the packing box in case that this
device needs re-packing.
14. Open the packing box and take out the patient circuit and bag arm assembly.
15. Install the patient circuit on one side of the device. Take care to align the assembly with
the circuit port and then push it to this device directly until it is stuck at the proper
position.
16. Open the small box containing bellows assembly and bellows housing gently. Remove
the plastic bag from the assembly and put the foam and plastic bag into the box.
17. Install the bellows onto the patient circuit. After installation, make sure that the bellows
are fully extended and surround the tabs of patient circuit.