6-16
14. Flow sensor calibration (refer to 7.3.2 Flow calibration (Service)).
15. O2 sensor calibration (refer to 7.3.6 O2 Sensor Calibration).
6.5 Other Materials Need to be Checked and Replaced
6.5.1 Maintenance and Replacing of Battery
It is no need to maintain the lithium battery. If the battery cannot work normally, replace it as
follows:
1. Open the access door of the anesthesia machine.
2. Open the battery box.
3. Remove the old battery.
4. Install new battery.
5. Closed the battery box.
6. Closed the rear cover plate.
Please only adopt the battery (P/N: 022-000008-00) which Mindray anesthesia machine supports.
6.6 Tests after Maintenance
After the anesthesia machine at the client end is maintained, some routine tests are required to
check if the current status of the anesthesia machine is normal. The following table lists the
routine tests.
SN
Test item
Functional description
Test interval
1
Check the
mechanical
ventilation
mode
1. Check if mechanical ventilation is provided normally
and if an alarm occurs.
2. Check if the preset values of pressure and TV are
same to the measured values.
3. Check if the pressure measured by the pressure
sensor is same to that indicated by the airway pressure
gauge and if the TV measured by the flow sensor is
same to that indicated by the graduation on the bellows
housing.
4. Roughly judge if the breathing system has a
significant leak by observing how much fresh gas is
compensatedt and observing if the folding bag
collapses.
After each
service or at
the time of
return visit
2
Breathing
system leak
test in
1. Check the pneumatic circuit in mechanical ventilation
mode for leaks, including bellows, drive gas circuit,
sodalime canister, patient tubes, flow sensors and their
After each
service or at
the time of
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...